Computerized Imaging Reference Systems, Incorporated (CIRS) 公司在1983年以改進(jìn)現(xiàn)有組織模擬方法和計(jì)算機(jī)斷層掃描提供定量的參考標(biāo)準(zhǔn)為目標(biāo)成立了。如今,CIRS被認(rèn)為是全球領(lǐng)先的等效幻影,放療,醫(yī)學(xué)成像模擬器以及程序培訓(xùn)...[查看更多]
For all computed tomography systems, the Food and Drug Administration recommends measuring the CT Dose Index (CTDI). Each section of the CIRS CT Dose Phantoms can provide separate dose information. The user can also measure maximum, minimum and mid-range values of the nominal tomographic section thickness when performing dose profile measurements.
Each phantom consists of a set of nesting 15 cm thick solid PMMA disks measuring 16 cm (head) and 32 cm (body) in diameter. The adult head disk is also suitable for pediatric body measurements. The Model 007A includes a third nesting disk measuring 10 cm in diameter for pediatric head measurements. Handles on the body and head are provided for ease in handling and maneuverability.
Through holes measuring 1.31 cm in diameter will accommodate standard CT probes. Acrylic rods are provided to plug the holes when not in use. The acrylic rods are machined to receive 1 mm diameter TLD rods.
The Model 007 & 007A are manufactured to comply with the FDA’s performance standard, 21 CFR 1020.33 that details the measurement requirements.
Features:
?Abdominal, Adult Head, Pediatric Head configurations
?PMMA disks and plugs with density of 1.19 g/cc
?13.1 mm diameter holes sized for standard CT Dose probes
?Nesting PMMA disks minimize storage space
Compatible with all CT scanners:
An optional Support Bracket, Model 007-01, can be used to suspend the CT Dose Phantoms above the imaging couch and align it along the axis of X-ray tube rotation. This enables the phantom to be used to assess CT dose in helical mode or any mode that requires the extended travel of the imaging couch or a wide beam. This set-up might be used to address the dosimetry approach described in TG111. An additional application of the support bracket is to provide a body to simulate continous scatter radiation from the patient during helical CT for dose safety measurements inside and outside the exam room.